ONLINE MEASUREMENT OF BRAIN GLUTAMATE WITH AN ENZYME/POLYMER-COATED TUBULAR ELECTRODE
ONLINE MEASUREMENT OF BRAIN GLUTAMATE WITH AN ENZYME/POLYMER-COATED TUBULAR ELECTRODE
复制标题
DOI:
10.1021/ac00085a016
复制
发表时间:
1994-07-01
影响因子:
7.4
通讯作者:
FILLENZ, M
中科院分区:
文献类型:
--
作者:
BERNERS, MOM;BOUTELLE, MG;FILLENZ, M
An on-line tubular electrode system is described which provides real time measurements of glutamate in brain microdialysate. It is based on the enzyme glutamate oxidase (EC 1.4.3.11) and the detection of H2O2 on a platinum electrode at 600 mV vs Ag/AgCI. The enzyme is immobilized in a layer of o-phenylenediamine electropolymerized at 750 mV in a phosphate buffer, pH 7.0. The layer is 10-15 nm thick and enables good enzyme loading and fast response time. The ability of the polymer to block out electroactive interferents like ascorbate and uric acid combined with a preoxidation system run at 600 mV resulted in a virtually interference free glutamate assay with a lower detection limit of 0.3 mu M in the presence of physiological levels of interferents, a sensitivity of 4 nA/mu M, and a linear region up to 30 mu M. The system is optimized for use with a microdialysis probe implanted in the brain and perfused at 2 mu L/min. It provides a time resolution of