Correction due to aperture in transmission interference microscopes
Correction due to aperture in transmission interference microscopes
复制标题
由于透射干涉显微镜的孔径而进行的校正
DOI:
--
复制
发表时间:
1958
期刊:
影响因子:
--
通讯作者:
L. Johansson
中科院分区:
文献类型:
--
作者:
E. Ingelstam;L. Johansson
The fringe spacing in transmission interference microscopes is shown to require a correction due to the oblique rays, with respect to the ordinary value valid for perpendicular rays. The spacing, and accordingly the measured path differences, t(nx - n0), are smaller than the normal values (opposite to the case of reflexion microscopes). Calculations are made for different values of condenser N.A. and varied parameters, showing corrections up to - 15%. The fringe intensities also decrease with the order of interference when it is low. Experiments confirm these findings. The effect is discussed in relation to some general ideas of (x, y)-fidelity and (z)-fidelity as obtained in a limited optical information channel.