A whole-blood flow cytometric assay for leukocyte CD11b expression using fluorescence signal triggering

A whole-blood flow cytometric assay for leukocyte CD11b expression using fluorescence signal triggering
复制标题

DOI:
10.1034/j.1600-0609.2000.90198.x
复制
发表时间:
2000-07-01
影响因子:
3.1
通讯作者:
Hjemdahl, P
Hjemdahl, P
中科院分区:
医学3区
文献类型:
--
作者:
Li, NL;Halldén, G;Hjemdahl, P

文献摘要

被引文献

相似文献

已开发并评价了用于测量全血中白细胞CD 11b表达的流式细胞术测定。该方法是基于荧光泛白细胞标志物RPE-CD 45触发流式细胞仪。这使得能够在全血中进行流式细胞术分析,并避免与细胞纯化和溶血相关的体外伪影。我们的方法学评价建议采用以下常规程序:用柠檬酸钠作为抗凝剂取样,在22 ℃下孵育样品,并用0.5%甲醛盐水温和固定样品。后者在24 h内提供了良好的样品稳定性。此外,该测定提供了良好的测定再现性、低标记抗体消耗和最小的样品操作(
A flow cytometric assay for measurements of leukocyte CD11b expression in whole blood has been developed and evaluated. The method is based on triggering of the flow cytometer by a fluorescent pan leukocyte marker, RPE-CD45. This enabled flow cytometric analysis in whole blood, and avoidance of in vitro artefacts related to cell purification and hemolysis. Our methodological evaluation suggested the following routine procedure: sampling with sodium citrate as the anticoagulant, sample incubation at 22 degrees C, and mild sample fixation with 0.5% formaldehyde saline. The latter provided good sample stability during 24 h. Moreover, the assay provided good assay reproducibility, low labelling antibody consumption, and minimal sample manipulation (