A low-cost photoacoustic microscopy system with a laser diode excitation
A low-cost photoacoustic microscopy system with a laser diode excitation
复制标题
DOI:
10.1364/boe.5.003053
复制
发表时间:
2014-09-01
影响因子:
3.4
通讯作者:
Zhu, Quing
中科院分区:
文献类型:
--
作者:
Wang, Tianheng;Nandy, Sreyankar;Zhu, Quing
Photoacoustic microscopy (PAM) is capable of mapping microvasculature networks in biological tissue and has demonstrated great potential for biomedical applications. However, the clinical application of the PAM system is limited due to the use of bulky and expensive pulsed laser sources. In this paper, a low-cost optical-resolution PAM system with a pulsed laser diode excitation has been introduced. The lateral resolution of this PAM system was estimated to be 7 mu m by imaging a carbon fiber. The phantoms made of polyethylene tubes filled with blood and a mouse ear were imaged to demonstrate the feasibility of this PAM system for imaging biological tissues. (C) 2014 Optical Society of America