Measurement of stresses in three orthogonal directions at the residual limb-prosthetic socket interface
Measurement of stresses in three orthogonal directions at the residual limb-prosthetic socket interface
复制标题
残肢-假肢接受腔界面三个正交方向应力的测量
DOI:
--
复制
发表时间:
1993
期刊:
影响因子:
--
通讯作者:
C. Daly
中科院分区:
文献类型:
--
作者:
J. Sanders;C. Daly
Transducers were developed to measure stresses in three orthogonal directions applied on a below-knee amputee residual limb within a prosthetic socket during walking. Strain-gage measurement techniques were used for all three orthogonal directions. Instrumentation error for four transducers averaged less than 4.2% full-scale output (FSO) for the normal direction and less than 0.9% FSO for shear directions. Measurements from amputee subjects collected during walking trials are presented. >