Brain SPECT Simulation Using Half-Cone-Beam Collimation and Single-Revolution Helical-Path Acquisition.
Brain SPECT Simulation Using Half-Cone-Beam Collimation and Single-Revolution Helical-Path Acquisition.
复制标题
使用半锥束准直和单转螺旋路径采集进行脑部 SPECT 模拟。
DOI:
10.1109/tns.2007.897826
复制
发表时间:
2007
影响因子:
1.8
通讯作者:
Metzler,ScottD
中科院分区:
文献类型:
--
作者:
Ter-Antonyan,Ruben;Jaszczak,RonaldJ;Bowsher,JamesE;Greer,KimL;Metzler,ScottD
In this study related to human brain SPECT imaging, simulation of half-cone-beam collimation and helical-path data acquisition is performed. We discuss problems related to circular-orbit acquisition using cone-beam collimation, such as shoulder interference resulting in object truncation, and insufficient sampling of the object resulting in axial distortions in the reconstructed images. We demonstrate that a triple-camera SPECT system with half-cone-beam collimation and single-revolution helical-path acquisition eliminates both issues and offers substantially improved sampling and almost artifact-free reconstruction of the object.