Lateral resolution enhancement of laser scanning microscopy by a higher-order radially polarized mode beam
Lateral resolution enhancement of laser scanning microscopy by a higher-order radially polarized mode beam
复制标题
DOI:
10.1364/oe.19.015947
复制
发表时间:
2011-08-15
期刊:
影响因子:
3.8
通讯作者:
Sato, Shunichi
中科院分区:
文献类型:
--
作者:
Kozawa, Yuichi;Hibi, Terumasa;Sato, Shunichi
We demonstrate that the lateral resolution of confocal laser scanning microscopy is dramatically improved by a higher-order radially polarized (HRP) beam with six concentric rings. This beam was generated simply by inserting liquid crystal devices in front of an objective lens. An HRP beam visualized aggregated 0.17 mu m beads individually and is also applicable to biological imaging. This method can extend the capability of conventional laser scanning microscopes without modification of the system, with the exception of the addition of the liquid crystal devices in the optical path. (C)2011 Optical Society of America