Preliminary Evaluations of a Self-Contained Anthropomorphic Transfemoral Prosthesis.

Preliminary Evaluations of a Self-Contained Anthropomorphic Transfemoral Prosthesis.
复制标题

DOI:
10.1109/tmech.2009.2032688
复制
发表时间:
2009
期刊:
IEEE/ASME transactions on mechatronics : a joint publication of the IEEE Industrial Electronics Society and the ASME Dynamic Systems and Control Division
影响因子:
--
通讯作者:
Goldfarb M
Goldfarb M
中科院分区:
其他
文献类型:
--
作者:
Sup F;Varol HA;Mitchell J;Withrow TJ;Goldfarb M

文献摘要

被引文献

相似文献

This paper presents a self-contained powered knee and ankle prosthesis, intended to enhance the mobility of transfemoral amputees. A finite-state based impedance control approach, previously developed by the authors, is used for the control of the prosthesis during walking and standing. Experiments on an amputee subject for level treadmill and overground walking are described. Knee and ankle joint angle, torque, and power data taken during walking experiments at various speeds demonstrate the ability of the prosthesis to provide a functional gait that is representative of normal gait biomechanics. Measurements from the battery during level overground walking indicate that the self-contained device can provide more than 4500 strides, or 9 km, of walking at a speed of 5.1 km/h between battery charges.