Development of a Mouse Biomagnetic Measurement System by Using a High-TC SQUID Magnetometer
Development of a Mouse Biomagnetic Measurement System by Using a High-TC SQUID Magnetometer
复制标题
DOI:
10.1109/tasc.2010.2094993
复制
发表时间:
2011-06-01
影响因子:
1.8
通讯作者:
Lee, Yong-Ho
中科院分区:
文献类型:
--
作者:
Kim, In-Seon;Ahn, San;Lee, Yong-Ho
We have developed a biomagnetic measurement system for mouse by using high-T-C SQUID magnetometer. The system consists of a high-T-C SQUID magnetometer, a table-top magnetically shielded box, and a nonmagnetic dewar with a warm-to-cool distance of about 4 mm. Mice with weights of about 27 g were anesthetized and fixed in supine position on a plastic xyz-stage inside the shield box. The SQUID readout signal was digitized in 24-bit and recorded with sampling rate of 1280 Hz. We carried out a scanning measurement by changing x-and y-position of the subject inside the shield box. Finally, we could obtain clear mouse MCG waveforms of about 20 pT in peak amplitude with well defined P-, QRS-, and T-wave. We demonstrated the high-T-C SQUID magnetometer based biomagnetic measurement system is useful for monitoring mouse MCG signals with sufficient noise-to-signal ratio.