Capillary array isoelectric focusing with laser-induced fluorescence detection: milli-pH unit resolution and yoctomole mass detection limits in a 32-channel system.

Capillary array isoelectric focusing with laser-induced fluorescence detection: milli-pH unit resolution and yoctomole mass detection limits in a 32-channel system.
复制标题

DOI:
10.1007/s00216-010-3595-x
复制
发表时间:
2010-08
影响因子:
4.3
通讯作者:
Dovichi, Norman J.
Dovichi, Norman J.
中科院分区:
化学2区
文献类型:
--
作者:
Dada, Oluwatosin O.;Ramsay, Lauren M.;Dickerson, Jane A.;Cermak, Nathan;Jiang, Rong;Zhu, Cuiru;Dovichi, Norman J.

文献摘要

参考文献

被引文献

相似文献

我们报告了一个多重毛细管电泳系统采用了32毛细管阵列与微机械鞘流池作为检测室。用473 nm激光束同时激发样品流,并将荧光发射在具有一对双重消色差透镜的CCD照相机上成像。该仪器对荧光素的区带电泳质量检测限为380 ± 120 μ mol。荧光标准品的毛细管等电聚焦产生平均宽度为0.0029 ± 0.0008 pH的峰。毛细管涂层稳定性限制了分析的重现性。
We report a multiplexed capillary electrophoresis system employing an array of 32 capillaries with a micromachined sheath flow cuvette as the detection chamber. The sample streams were simultaneously excited with a 473 nm laser beam, and the fluorescence emission was imaged on a CCD camera with a pair of doublet achromat lens. The instrument produced mass detection limits of 380 ± 120 yoctomoles for fluorescein in zone electrophoresis. Capillary isoelectric focusing of fluorescent standards produced peaks with an average width of 0.0029 ± 0.0008 pH. Capillary coating stability limits the reproducibility of the analysis.
DOI: 10.1021/ac001001c
发表时间: 2001-03-15
影响因子: 7.4
作者:
Zhang, JZ;Yang, MJ;Dovichi, NJ
通讯作者: Dovichi, NJ
DOI: 10.1093/nar/20.18.4873
发表时间: 1992-09-25
影响因子: 14.9
作者:
CHEN, DY;HARKE, HR;DOVICHI, NJ
通讯作者: DOVICHI, NJ
DOI: 10.1021/ac8025948
发表时间: 2009-03-01
影响因子: 7.4
作者:
Ramsay, Lauren M.;Dickerson, Jane A.;Dada, Oluwatosin;Dovichi, Norman J.
通讯作者: Dovichi, Norman J.
DOI: 10.1002/anie.200460508
发表时间: 2004-01-01
影响因子: 16.6
作者:
Wetzl, BK;Yarmoluk, SM;Wolfbeis, OS
通讯作者: Wolfbeis, OS
DOI: 10.1002/elps.1150140172
发表时间: 1993-05-01
期刊: ELECTROPHORESIS
影响因子: 2.9
作者:
WU, JQ;PAWLISZYN, J
通讯作者: PAWLISZYN, J