An instrument for nitric oxide measurements in the stratosphere.

An instrument for nitric oxide measurements in the stratosphere.
复制标题

DOI:
10.1063/1.1686726
复制
发表时间:
1974-06
期刊:
The Review of scientific instruments
影响因子:
--
通讯作者:
B. Ridley;L. Howlett
B. Ridley;L. Howlett
中科院分区:
其他
文献类型:
--
作者:
B. Ridley;L. Howlett

文献摘要

被引文献

相似文献

本文介绍了一种全自动平流层NO原位测量装置。在21.3 km处的典型响应度为1860计数/秒ppbv。对于1秒的测量时间常数,由噪声确定的检测限为0.03 ppbv。该仪器已从气球平台飞行到30.8公里,从飞机平台飞行到12.2公里。18.3公里。
A completely automatic chemiluminescent instrument has been developed for in situ measurements of NO in the stratosphere. Signal intensity is linear in NO. Typical responsivity at 21.3 km is 1860 counts/sec ppbv. With a 1 sec measuring time constant, the detection limit as determined by noise is 0.03 ppbv. The instrument has been flown from balloon platforms to 30.8 km and from aircraft platforms between 12.2. and 18.3 km.