Liquid Core Waveguide Spectrophotometry for the Sensitive Determination of Nitrite in River Water Samples
Liquid Core Waveguide Spectrophotometry for the Sensitive Determination of Nitrite in River Water Samples
复制标题
DOI:
10.2116/analsci.22.1017
复制
发表时间:
2006-07
影响因子:
1.6
通讯作者:
H. Takiguchi;A. Tsubata;M. Miyata;T. Odake;H. Hotta;T. Umemura;K. Tsunoda
中科院分区:
文献类型:
--
作者:
H. Takiguchi;A. Tsubata;M. Miyata;T. Odake;H. Hotta;T. Umemura;K. Tsunoda
A flow injection analysis system was built with a liquid core waveguide spectrophotometric detector using an 80 cm Teflon AF-1600 capillary tube (2,2-bistrifluoromethyl-4,5-difluoro-1,3-dioxole/tetrafluoroethylene). The system was applied to determine nitrite ion in river water samples. The lower limit of detection for nitrite was 2.1 nmol dm^–3 (0.1 ng dm^–3 as NO_2^–) and the relative standard deviation of measurements was typically 0.56% ( n = 5) at 0.21 μmol dm^–3.