Liquid Core Waveguide Spectrophotometry for the Sensitive Determination of Nitrite in River Water Samples

Liquid Core Waveguide Spectrophotometry for the Sensitive Determination of Nitrite in River Water Samples
复制标题

DOI:
10.2116/analsci.22.1017
复制
发表时间:
2006-07
影响因子:
1.6
通讯作者:
H. Takiguchi;A. Tsubata;M. Miyata;T. Odake;H. Hotta;T. Umemura;K. Tsunoda
H. Takiguchi;A. Tsubata;M. Miyata;T. Odake;H. Hotta;T. Umemura;K. Tsunoda
中科院分区:
化学4区
文献类型:
--
作者:
H. Takiguchi;A. Tsubata;M. Miyata;T. Odake;H. Hotta;T. Umemura;K. Tsunoda

文献摘要

被引文献

相似文献

采用80 cm Teflon AF-1600毛细管(2,2-双三氟甲基-4,5-二氟-1,3-间二氧杂环戊烯/间二氧杂环戊烯),以液芯波导分光光度检测器建立了流动注射分析系统。将该系统应用于河水样品中亚硝酸根离子的测定。亚硝酸盐的检测下限为2.1 nmol dm^-3(以NO_2^-计为0.1 ng dm^-3),在0.21 μmol dm^-3时,测量的相对标准偏差通常为0.56%(n = 5)。
A flow injection analysis system was built with a liquid core waveguide spectrophotometric detector using an 80 cm Teflon AF-1600 capillary tube (2,2-bistrifluoromethyl-4,5-difluoro-1,3-dioxole/tetrafluoroethylene). The system was applied to determine nitrite ion in river water samples. The lower limit of detection for nitrite was 2.1 nmol dm^–3 (0.1 ng dm^–3 as NO_2^–) and the relative standard deviation of measurements was typically 0.56% ( n = 5) at 0.21 μmol dm^–3.