iTRAQ reagent-based quantitative proteomic analysis on a linear ion trap mass spectrometer

iTRAQ reagent-based quantitative proteomic analysis on a linear ion trap mass spectrometer
复制标题

DOI:
10.1021/pr070291b
复制
发表时间:
2007-11-01
影响因子:
4.4
通讯作者:
Higgins, LeeAnn
Higgins, LeeAnn
中科院分区:
生物学2区
文献类型:
--
作者:
Griffin, Timothy J.;Xie, Hongwei;Higgins, LeeAnn

文献摘要

被引文献

相似文献

For proteomic analysis using tandem mass spectrometry, linear ion trap instruments provide unsurpassed sensitivity but unreliably detect low mass peptide fragments, precluding their use with iTRAQ reagent-labeled samples. Although the popular LTQ linear ion trap supports analyzing iTRAQ reagent-labeled peptides via pulsed Q dissociation, PQD, its effectiveness remains questionable. Using a standard mixture, we found careful tuning of relative collision energy necessary for fragmenting iTRAQ reagent-labeled peptides, and increasing microscan acquisition and repeat count improves quantification but identifies somewhat fewer peptides. We developed software to calculate abundance ratios via summing reporter ion intensities across spectra matching to each protein, thereby providing maximized accuracy. Testing found that results closely corresponded between analysis using optimized LTQ-PQD settings plus our software and using a Qstar instrument. Thus, we demonstrate the effectiveness of LTQ-PQD analyzing iTRAQ reagent-labeled peptides, and provide guidelines for successful quantitative proteomic studies.