Comparison of xenon triple point realizations
Comparison of xenon triple point realizations
复制标题
氙三相点实现的比较
DOI:
10.1088/1681-7575/aaee3a
复制
发表时间:
2018
期刊:
影响因子:
2.4
通讯作者:
D. Giraudi
中科院分区:
文献类型:
--
作者:
P. Steur;P. Rourke;D. Giraudi
It is widely recognized that the Mercury triple point (MTP) being situated very close to the Water triple point (WTP) constitutes a weakness in the International Temperature Scale of 1990 (ITS-90), in addition to safety concerns related to the use and transportation of Mercury. As such, a substitution for a safer, high-quality fixed point about half way between the Argon and Water triple points would be highly desirable. Now, a direct comparison is described of a Xenon cell filled in 2005 by the National Research Council Canada (NRC) and a more recently produced cell of the Istituto Nazionale di Ricerca Metrologica (INRIM). The present paper discusses the INRiM 2017 measurements on both the INRiM and NRC cells, with a follow-up measurement at NRC, and presents the difference between the two cells, (0.17 ± 0.08) mK with the uncertainties of each cell’s realization of the Xenon triple point (XeTP), 0.11 mK for the INRiM cell and 0.07 mK for the NRC cell (k = 1). In addition, the effect of substituting Mercury with Xenon on Type 1 non-uniqueness (‘SRI’, subrange inconsistency), Type 3 non-uniqueness (‘NU3’, cSPRT variability) and propagation of fixed point realization uncertainty is shown and discussed.