Materials and methods for higher performance screen-printed flexible MRI receive coils.
Materials and methods for higher performance screen-printed flexible MRI receive coils.
复制标题
DOI:
10.1002/mrm.26399
复制
发表时间:
2017-08
影响因子:
3.3
通讯作者:
Arias AC
中科院分区:
文献类型:
--
作者:
Corea JR;Lechene PB;Lustig M;Arias AC
To develop methods for characterizing materials used in screen-printed MRI coils and improve SNR with new lower-loss materials. An experimental apparatus is created to characterize dielectric properties of plastic substrates used in receive coils. Coils are fabricated by screen printing conductive ink onto several plastic substrates. Unloaded and sample loaded quality factor (QUnloaded/QLoaded) measurements and scans on a 3T scanner are used to characterize coil performance. An experimental method is developed to describe the relationship between a coil's QUnloaded and the SNR it provides in images of a phantom. Additionally, 3T scans of a phantom and the head of a volunteer are obtained with a proof-of-concept printed 8-channel array and the results are compared to a commercial 12-channel array. Printed coils with optimized substrates can exhibit up to 97% of the image SNR when compared to a traditional coil on a loading phantom. QUnloaded and the SNR of coils are successfully correlated. The printed array results in images comparable to the quality given by the commercial array. With the methods and materials proposed, the SNR of printed coils approaches that of commercial coils while using a new fabrication technique that provides more flexibility and close contact to the patient's body.