Comparison of two spatially resolved near-infrared photometers in the detection of tissue oxygen saturation: poor reliability at very low oxygen saturation.
Comparison of two spatially resolved near-infrared photometers in the detection of tissue oxygen saturation: poor reliability at very low oxygen saturation.
复制标题
两种空间分辨近红外光度计在组织氧饱和度检测中的比较:氧饱和度极低时可靠性较差。
DOI:
10.1042/cs1010715
复制
发表时间:
2001
期刊:
影响因子:
6
通讯作者:
M. Ferrari
中科院分区:
文献类型:
--
作者:
T. Komiyama;V. Quaresima;H. Shigematsu;M. Ferrari
Two spatially resolved oximeters, NIRO-300 and OM-200, were compared with regard to the measurement of oxygen saturation values in two forearm muscle groups at rest and during arterial occlusion in nine healthy volunteers. There was a significant correlation between the muscle oxygen saturation values obtained at rest using the two oximeters (n=33, r(2)=0.43, P<0.0001), whereas these values were significantly different during arterial occlusion. Thus, although there was good agreement between muscle oxygen saturation values measured using the two oximeters, the operating range of the tissue oximeters should be recognized and indicated.