Decoupling Design of Z2 Superconducting Shim Coils for 9.4-T MRI Superconducting Magnet
Decoupling Design of Z2 Superconducting Shim Coils for 9.4-T MRI Superconducting Magnet
复制标题
DOI:
10.1109/tasc.2016.2530310
复制
发表时间:
2016-03
影响因子:
1.8
通讯作者:
Yi Li;Qiuliang Wang;Xuchen Zhu;Z. Ni;Lei Wang
中科院分区:
文献类型:
--
作者:
Yi Li;Qiuliang Wang;Xuchen Zhu;Z. Ni;Lei Wang
A set of superconducting shim coils for the 9.4-T whole-body MRI superconducting magnet have been designed. A minimum criterion for the Z2 shim coils' decoupling from the magnet is introduced for the safety of the Z2 shim coils during the magnet quench. The traditional two-pair structure of the Z2 shim coils is introduced based on analytical method. A new five-coil structure of the Z2 shim coils for decoupling design is proposed. Optimized solutions with respect to different design parameters are obtained and analyzed. Solution is chosen for the magnet in consideration of the capacity that the magnet needs and the field error that other shim methods can deal with.