Imaging cardiac activity by the D-bar method for electrical impedance tomography

Imaging cardiac activity by the D-bar method for electrical impedance tomography
复制标题

DOI:
10.1088/0967-3334/27/5/s04
复制
发表时间:
2006-05-01
影响因子:
3.2
通讯作者:
Siltanen, S.
Siltanen, S.
中科院分区:
工程技术3区
文献类型:
--
作者:
Isaacson, D.;Mueller, J. L.;Siltanen, S.

文献摘要

被引文献

相似文献

描述了一种实用的D-bar算法,用于从二维几何结构中电极上的EIT数据重建电导率变化。该算法基于Nachman(1996)的全局唯一性证明。数学。143 71-96)的二维逆电导率问题。结果显示了从放置在人体胸部周围的电极上收集的数据重建的结果,以重建躯干的二维截面。图像显示心脏周期中电导率的变化。
A practical D-bar algorithm for reconstructing conductivity changes from EIT data taken on electrodes in a 2D geometry is described. The algorithm is based on the global uniqueness proof of Nachman (1996 Ann. Math. 143 71-96) for the 2D inverse conductivity problem. Results are shown for reconstructions from data collected on electrodes placed around the circumference of a human chest to reconstruct a 2D cross-section of the torso. The images show changes in conductivity during a cardiac cycle.