Resolution Properties Of A Computed Radiographic System
Resolution Properties Of A Computed Radiographic System
复制标题
计算机射线照相系统的分辨率特性
DOI:
10.1117/12.953211
复制
发表时间:
1989
期刊:
影响因子:
--
通讯作者:
T. Fujikawa
中科院分区:
文献类型:
--
作者:
H. Fujita;J. Morishita;K. Ueda;D. Tsai;A. Ohtsuka;T. Fujikawa
The analysis of spatial-resolution properties in terms of the modulation transfer function (MTF) has been presented in a computed radiographic (CR) system (FCR-101) with the photostimulable-phosphor plate (imaging plate, IP). The newly devised method of determining the presampling MTF which includes the x-ray detector (IP) unsharpness and the unsharpness of the sampling aperture is described in which an image of a slightly-angulated lead slit relative to a horizontal or vertical direction is employed. The IP MTFs as an analog MTF in the system, the presampling MTFs for different types of IPs, different sampling distances, different versions of IPs, simultaneous multisection tomography and magnification radiography, and the laser-printer MTFs as display MTF are measured and shown. The effective sampling aperture MTFs calculated indicate that the noticeable degradation of resolution occurs at the stage of image data sampling. The usefulness of the magnification technique for mammography and bone radiography is demonstrated. It is shown that both of the digital MTF and the overall MTF are difficult to use for general purpose due to the aliasing artifacts. The effect of glare on the contrast is also characterized by lead-disk method. The glare fraction is found to be approximately 6.5%.
影响因子:
3.8
作者:
Giger,ML;Doi,K;Metz,CE
通讯作者:
Metz,CE
影响因子:
3.8
作者:
Fujita,H;Doi,K
通讯作者:
Doi,K
影响因子:
3.8
作者:
Fujita,H;Doi,K;Giger,ML;Chan,HP
通讯作者:
Chan,HP