Acoustic response from adherent targeted contrast agents.
Acoustic response from adherent targeted contrast agents.
复制标题
DOI:
10.1121/1.2364303
复制
发表时间:
2006-12
期刊:
影响因子:
--
通讯作者:
Shukui Zhao;D. Kruse;K. Ferrara;P. Dayton
中科院分区:
文献类型:
--
作者:
Shukui Zhao;D. Kruse;K. Ferrara;P. Dayton
In ultrasonic molecular imaging, encapsulated micron-sized gas bubbles are tethered to a blood vessel wall by targeting ligands. A challenging problem is to detect the echoes from adherent microbubbles and distinguish them from echoes from nonadherent agents and tissue. Echoes from adherent contrast agents are observed to include a high amplitude at the fundamental frequency, and significantly different spectral shape compared with free agents (p <0.0003). Mechanisms for the observed acoustical difference and potential techniques to utilize these differences for molecular imaging are proposed.