Forceps through a 14F flexible sheath for safe retrieval of a dislodged left atrial appendage occluder

Forceps through a 14F flexible sheath for safe retrieval of a dislodged left atrial appendage occluder
复制标题

DOI:
10.1111/jce.14691
复制
发表时间:
2020
影响因子:
2.7
通讯作者:
Zhong Jingquan
Zhong Jingquan
中科院分区:
医学3区
文献类型:
--
作者:
Qiu Libin;Rong Bing;Zhang Kai;Zhong Jingquan

文献摘要

相似文献

The LAmbre occluder (Lifetech Scientific, Shenzhen, China), a new device for left atrial appendage occlusion (LAAO), is increasingly used, but the retrieval procedure for dislodgements has rarely been reported in humans. We herein reported a case of safe retrieval of dislodged LAmbre occluder. The occluder dislodged to the left atrium (LA) at the end of the LAAO procedure. We failed to retrieve the device in the LA using a LASSO catheter and forceps through an 8.5F sheath. After it suddenly flowed into the aortic arch, we successfully retrieved it using the same forceps through a 14F flexible sheath. We also discuss the reasons for device dislodgment and experiences of device retrieval. We think that forceps can be used to retrieve an embolized LAmbre device in the aorta through a 14F flexible sheath.