课题基金 / 基金详情

The development of the apparatus for structure strength test of cervical vertebra

The development of the apparatus for structure strength test of cervical vertebra
颈椎结构强度测试仪的研制
批准号:
16500292
负责人:
KAWANO Shunichi
金额:
$1.02万
依托单位:
依托单位国家:
日本
项目类别:
Grant-in-Aid for Scientific Research (C)
财政年份:
2004
资助国家:
日本
项目状态:
已结题
起止时间:
2004 至 2005

项目摘要

项目成果

相关文献

中文摘要
翻译
点击翻译按钮获取中文摘要
英文摘要
In this research, we describe an apparatus for the structure strength test of cervicalp vertebra and investigate the mechanical behavior of cervicalp vertebra (C.V) by using this appratus. The test apparatus is composed of 4 linear guide actuators, two AC servo motors, 4 load and 2 torque sensors. In order to obtain the pure bending moment of C.V, the reaction loads caused by the torque are canceled by the controllers. Bend tests are carried out on a cervicalp vertebra of pig.The following conclusions are obtained in this research.1)Bending of C.V ; The bending torques decrease with an decreasing reaction load and the pure bending moments of C.V are obtained by using the test apparatus.2)Compression of C.V ; The compression load on the C.V brings about the large reaction load on the vertical axis of a longitudinal C.V. The experimental result shows that the C.V is buckled easiy by the bending torque on C.V conbined with the compression load.