中文摘要
用于全髋关节置换术的超高分子量聚乙烯(UHMWPE)衬垫。近年来,为了提高运动范围,衬垫厚度变薄、大直径也在不断发展。本研究采用3.5/4.0mm厚度44mm直径的PE衬垫,在标准模拟器测试条件下进行了500万次的髋关节模拟器测试。与未掺杂PE衬垫相比,0.1% α -生育酚(0.1%重量比)掺杂杯没有磨损优势。
英文摘要
The ultra highly molecular weight polyethylene (UHMWPE) liner used in total hip replacement. Recently, thinning of thickness and large-diameter of liner is progressing for the range of motion improved. In this study, hip simulator test had run by using 3.5/4.0mm thickness 44mm diameter PE liner under standard simulator test conditionsup to 5 million cycles. Compared to non doped PE liner, There was no wear advantage related to 0.1% alpha-tocopheral (0.1% by weight) doped cup.