Platelet rich plasmas are tested using the Capture-P kit (Immucor).
Platelet monolayers in microtiter wells are incubated with anti-PL-
A1, washed, and exposed to anti-immunoglobulin G-coated red blood
cells. The coated red blood cells adhere to PL-A1 positive platelets
but not to PL-A1 negative platelets. Centrifugal sedimentation of
unattached red blood cells produces a central button, indicating
that platelets are PL-A1 negative. Of 520 donors, 15 (2.88%) tested
PL-A1 negative, which correlates well with the reported frequency of
PL-A2 in caucasians in the literature. All 15 PL-A1 negative donors
were confirmed by allele-specific restriction assays. Confirmatory
genotyping was not done on those donors testing PL-A1 positive. The
rate of PL-A1 negative donors correlates well with the published
rate for caucasians. The assay can be used to screen large numbers
of samples. Confirmatory testing can be reserved for sample testing
PL-A1 negative by solid phase. Ongoing work involves identifying the
rate of "false positives."