Assesment of a Chronic Subcutaneous Glucose Sensor
Assesment of a Chronic Subcutaneous Glucose Sensor
批准号:
6437813
负责人:
WILLIAM KENNETH WARD
金额:
$26.49万
依托单位:
依托单位国家:
美国
项目类别:
财政年份:
2002
资助国家:
美国
项目状态:
已结题
起止时间:
2002-04-01 至 2004-03-31
关键词:
angiogenesis bioengineering /biomedical engineering biomedical equipment development biosensor device blood glucose diabetes mellitus electrodes hypoglycemia immunocytochemistry laboratory rabbit medical implant science monitoring device statistics /biometry telemetry vascular endothelial growth factors
中文摘要
点击翻译按钮获取中文摘要
英文摘要
DESCRIPTION (provided by applicant): In the United States, diabetes mellitus is
the leading cause of end-stage kidney failure, blindness in adults under age
65, and is the second leading cause (after trauma) of limb amputation.
Poorly-controlled glucose levels are the major cause of these complications,
but tight glycemic control is difficult to safely achieve using present
technology. While a continuously-functioning glucose sensor would assist in
safely achieving tight glucose control, such devices are typified by
instability and loss of output over time due to the formation of foreign-body
scar tissue, which eventually surrounds the sensors. Miniaturized sensors could
be implanted under the skin with minimal discomfort. It is possible that the
surrounding scar capsule could be made much more "friendly" to a glucose sensor
by the slow release of growth factor compounds from the sensor surface. The
hypothesis is that such compounds would reduce the scar fibrosis and generate
many blood vessels in the capsule. These blood vessels are important to the
long-term function of the sensor given its need for continuous glucose and
oxygen delivery. A study will initially be performed in order to ascertain the
optimal dosage of the growth factor. Then the growth factor (or saline control)
will be slowly released over one month from miniaturized pumps into the tissue,
which directly contacts the sensor. The thickness of the capsule and the
formation of new blood vessels will measured by standard histologic staining
techniques and by endothelial Factor Vifi immunohistochemistry. Another major
problem of glucose sensors is short-term drift. It is frequently observed but
poorly understood. It now appears that it may be possible to obtain stable
readings from simultaneous recordings of multiple sensor electrodes. The
technique which will be used to separate the accurate electrode signals from
the inaccurate (outlying) signals is from the field of median statistics and is
called the ZMAD method of Rousseeauw. The ZMAD data processing will be
performed prospectively and continuously. In addition, the sensors and their
transmitters will be miniaturized with the help of a biotelemetry company,
MiniMitter. The body can be hostile to compounds which coat implanted devices.
We will compare two promising polyurethanes as sensor coats: a carbonate based-
vs. ether-based polymer. We will compare their long term function in studies
using rabbits, which will be also be used to ascertain the effect of the growth
factors and the real-time data processing.
期刊论文(0)
专著(0)
科研奖励(0)
会议论文
Assesment of a Chronic Subcutaneous Glucose Sensor
-
批准号:6621918
-
项目类别:
-
资助金额:$19.28万
-
财政年份:2002
-
负责人:WILLIAM KENNETH WARD
-
依托单位:
Passivating Proteins in Implantable Glucose Sensors
-
批准号:6663151
-
项目类别:
-
资助金额:$30.79万
-
财政年份:2002
-
负责人:WILLIAM KENNETH WARD
-
依托单位:
Passivating Proteins in Implantable Glucose Sensors
-
批准号:6792087
-
项目类别:
-
资助金额:$30.54万
-
财政年份:2002
-
负责人:WILLIAM KENNETH WARD
-
依托单位:
Passivating Proteins in Implantable Glucose Sensors
-
批准号:6588889
-
项目类别:
-
资助金额:$30.83万
-
财政年份:2002
-
负责人:WILLIAM KENNETH WARD
-
依托单位: