Theoretical evaluation of the high-resolution vacuum ultraviolet spectrometer on SURF II.
Theoretical evaluation of the high-resolution vacuum ultraviolet spectrometer on SURF II.
复制标题
SURF II 高分辨率真空紫外光谱仪的理论评估
DOI:
10.1364/ao.31.006724
复制
发表时间:
1992
期刊:
影响因子:
1.9
通讯作者:
H. M. Seyoum
中科院分区:
文献类型:
--
作者:
N. C. Das;R. Madden;W. Olson;H. M. Seyoum
The high-resolution vacuum-ultraviolet spectroscopic facility at SURF II, the Synchrotron Ultraviolet Radiation Facility at the National Institute of Standards and Technology in Gaithersburg, Md., consists of a fore-optics system of three cylindrical mirrors and a 6.65-m concave grating spectrometer using the off-plane Eagle mounting. To prepare for the evaluation of the actual performance of this nationalfacility spectrometer against theoretical expectations, we computed scanning parameters, spectral resolution, and the optimum curvature and tilt of both entrance and exit slits. It is planned eventually to replace the exit slit of this instrument with a two-dimensional array detector to increase data collection efficiency. Therefore a major motivation for this work is that the results on the tilt and curvature of the exit slit can be used to maximize the resolution obtainable with the array detector through data processing.