Intra-operative Real-Time 3-D Information Display System Based on Integral Videography
Intra-operative Real-Time 3-D Information Display System Based on Integral Videography
复制标题
基于一体化摄像的术中实时3D信息显示系统
DOI:
10.1007/3-540-45468-3_47
复制
发表时间:
2001
期刊:
影响因子:
--
通讯作者:
T. Dohi
中科院分区:
文献类型:
--
作者:
H. Liao;S. Nakajima;M. Iwahara;E. Kobayashi;I. Sakuma;N. Yahagi;T. Dohi
A real-time 3-D surgical navigation system that superimposes the real, intuitive 3-D image for medical diagnosis and operation was developed in this paper. This system creates 3-D image based on the principle of integral photography (IP), named “Integral Videography (IV)”, which can display geometrically accurate 3-D image and reproduce motion parallax without any need of special devices. 3-D image was superimposed on the surgical fields in the patient via a half-silvered mirror as if they could be seen through the body. In addition, a real-time IV algorithm for calculating the 3-D image of surgical instruments was used for registration between the location of surgical instruments and the organ during the operation. The experimental results of puncturing a point location and avoiding critical area showed the errors of this navigation system were in the range of 2~3mm. By introducing a display device with higher pixel density, accuracy of the system can be improved.