Analytical control of silica glass production. Voltammetric determination of titanium and iron in raw materials and silica glass samples
Analytical control of silica glass production. Voltammetric determination of titanium and iron in raw materials and silica glass samples
复制标题
石英玻璃生产的分析控制。
DOI:
--
复制
发表时间:
2000
期刊:
影响因子:
--
通讯作者:
J. Golimowski
中科院分区:
文献类型:
--
作者:
M. Gawryś;J. Golimowski
Abstract Adsorptive stripping voltammetric (AdSV) methods are presented for the determination of titanium and iron in quartz and silica glass samples obtained after pressure decomposition. Mandelic acid and catechol were used as complexing agents for titanium and iron, respectively. The method for titanium determination is based on the catalytic effect of chlorate ions. An insoluble residue remaining after decomposition of quartz and silica glass samples in HF+H2SO4 mixture was checked by energy-dispersive X-ray analyses. ET-AAS was applied as a reference method to AdSV measurements.