Influence of simulator kinematics on the wear of metal-on-metal hip prostheses
Influence of simulator kinematics on the wear of metal-on-metal hip prostheses
复制标题
DOI:
10.1243/0954411011533481
复制
发表时间:
2001-01-01
期刊:
影响因子:
1.8
通讯作者:
Fisher, J
中科院分区:
文献类型:
--
作者:
Firkins, PJ;Tipper, JL;Fisher, J
There is now considerable interest in metal-on-metal bearings for hip prostheses. Extremely low wear rates (0.1 mm(3)/10(6) cycles) have been reported in some simulator studies, while in vivo studies, although still very low, have shown wear rates of the order of 1 mm(3)/10(6) cycles. The aim of this study was to compare wear rates of metal-on-metal bearings in two hip simulators with different kinematic inputs. In the simulator with three independent input motions which produced an open elliptical wear path with a low level of eccentricity, the wear rates were very low as recorded previously in other simulators. In the simulator with two input motions which produced an open elliptical wear path with greater eccentricity the wear rate was at least ten times higher and closer to clinical values. The motion and kinematic conditions in the contact are critical determinants of wear in metal-on-metal bearings.