Clinical prototype of a plastic water-equivalent scintillating fiber dosimeter array for QA applications

Clinical prototype of a plastic water-equivalent scintillating fiber dosimeter array for QA applications
复制标题

DOI:
10.1118/1.2953564
复制
发表时间:
2008-08-01
期刊:
影响因子:
3.8
通讯作者:
Beaulieu, Luc
Beaulieu, Luc
中科院分区:
医学3区
文献类型:
--
作者:
Lacroix, Frederic;Archambault, Louis;Beaulieu, Luc

文献摘要

被引文献

相似文献

A clinical prototype of a scintillating fiber dosimeter array for quality assurance applications is presented. The array consists of a linear array of 29 plastic scintillation detectors embedded in a water-equivalent plastic sheet coupled to optical fibers used to guide optical photons to a charge coupled device (CCD) camera. The CCD is packaged in a light-tight, radiation-shielded housing designed for convenient transport. A custom designed connector is used to ensure reproducible mechanical positioning of the optical fibers relative to the CCD. Profile and depth dose characterization measurements are presented and show that the prototype provides excellent dose measurement reproducibility (+/- 0.8%) in-field and good accuracy (+/- 1.6% maximum deviation) relative to the dose measured with an IC10 ionization chamber.