PEMC sensor's mass change sensitivity is 20 pg/Hz under liquid immersion.
PEMC sensor's mass change sensitivity is 20 pg/Hz under liquid immersion.
复制标题
PEMC传感器在液浸状态下的质量变化灵敏度为20 pg/Hz。
DOI:
10.1016/j.bios.2005.11.020
复制
发表时间:
2006
影响因子:
12.6
通讯作者:
Mutharasan,Raj
中科院分区:
文献类型:
--
作者:
Campbell,GossettA;Mutharasan,Raj
To enhance the mass change sensitivity of the resonating piezoelectric-excited millimeter-sized cantilever (PEMC) sensors, we reduced its length and eliminated one layer of its composite structure. As a result the mass sensitivity of the second flexural mode increased by two orders of magnitude (from 10−9to 10−11g/Hz) and the resonant frequency increased by more than 5kHz. We demonstrate the effects of modification by detecting a model pathogen Group A Streptococcus (GAS) at 700cells/mL. The resonant frequency change of the second mode at concentrations of 700, 7×103, 7×105, 7×106, 7×107, and 7×109cells/mL resulted in, respectively, 3.1±0.5, 11.6±1, 15.7±1, 25.7±0.15, 28.5±2, and 40.5±3ng (n=3 for all) of pathogen attachment. A kinetic model for the binding is proposed and verified. The observed binding rate constant was found to be in the range of 0.051–0.166min−1. The significance of the results we report is that the modified PEMC sensors have high mass sensitivity that pathogens can be detected at very low concentration under liquid immersion conditions.