Transcatheter Valve Underexpansion Limits Leaflet Durability: Implications for Valve-in-Valve Procedures
Transcatheter Valve Underexpansion Limits Leaflet Durability: Implications for Valve-in-Valve Procedures
复制标题
DOI:
10.1007/s10439-016-1738-8
复制
发表时间:
2017-02-01
影响因子:
3.8
通讯作者:
Sun, Wei
中科院分区:
文献类型:
--
作者:
Martin, Caitlin;Sun, Wei
Transcatheter aortic valve (TAV) implantation within a failed bioprosthetic valve is a growing trend for high-risk patients. The non-compliant stent of the previous prosthesis may prevent full expansion of the TAV, which has been shown to distort the leaflet configuration, and has been hypothesized to adversely affect durability. In this study, TAV leaflet fatigue damage under cyclic pressurization in the setting of stent underexpansion by 0 (fully expanded), 1, 2 and 3 mm was simulated using finite element analysis to test this hypothesis. In the 2 and 3 mm underexpanded devices, the TAV leaflets exhibited severe pin-wheeling during valve closure, which increased leaflet stresses dramatically, and resulted in accelerated fatigue damage of the leaflets. The leaflet fatigue damage in the 1 mm underexpanded case was similar to that in the fully expanded case. Clinically a range of 10-15% underexpansion is generally considered acceptable; however, it was observed in this study that ae2 mm (ae9.1%) underexpansion, will significantly impact device durability. Further study is necessary to determine the impact of various deployment conditions, i.e. non-uniform and non-circular deployments and different implantation heights, on differing TAV devices, but it is clear that the normal TAV leaflet configuration must be preserved in order to preserve durability.