High-resolution x-ray spectra measured using tantalum superconducting tunnel junctions
High-resolution x-ray spectra measured using tantalum superconducting tunnel junctions
复制标题
使用钽超导隧道结测量的高分辨率 X 射线光谱
DOI:
10.1063/1.121603
复制
发表时间:
1998
影响因子:
4
通讯作者:
D. Goldie
中科院分区:
文献类型:
--
作者:
P. Verhoeve;N. Rando;A. Peacock;A. V. Dordrecht;B. Taylor;D. Goldie
The spectral response of a 100×100 μm2 tantalum based superconducting tunnel junction to 5.9 keV x-ray photons from a 55Fe source has been studied. In full illumination the energy resolution for the Mn Kα line complex is 56 eV, dominated by spatial nonuniformity in the response of the detector. When illuminating selectively a 5–10 μm diam spot in the center of the detector, the energy resolution improves to 22 eV, corresponding to 15.7 eV for the individual Mn Kα1 and Mn Kα2 lines. This exceeds the predicted theoretical energy resolution of 7.3 eV for this type of device by only a factor of ∼2.