A 20-Channel Grating Polychromator for Electron Cyclotron Emission Measurements on EAST
A 20-Channel Grating Polychromator for Electron Cyclotron Emission Measurements on EAST
复制标题
DOI:
10.13182/fst11-a11730
复制
发表时间:
2011-05
影响因子:
0.9
通讯作者:
Yong Liu;E. Li;B. Ling;A. Ti;Gary Taylor
中科院分区:
文献类型:
--
作者:
Yong Liu;E. Li;B. Ling;A. Ti;Gary Taylor
Abstract A 20-channel grating polychromator transferred from Princeton Plasma Physics Laboratory has been rebuilt for electron cyclotron emission measurements on EAST. This instrument measures the second-harmonic electron cyclotron emission from plasma with frequency range from 90 to 250 GHz, which corresponds to a central magnetic field (R0 = 1.7 m) of 2 to 3.5 T on EAST. The radial resolution of this instrument is [approximately]2.5 cm, and the poloidal spot size of the quasi-optic antenna is [approximately]3 cm. New preamplifiers are made and tested, based on the electronics of GPC-II on TFTR. These amplifiers have a gain of around 520, with a 400-kHz, 3-dB rolloff. Measurement results from the 2010 EAST experimental campaign show that the intensity of this instrument is [approximately]200 mV for electron temperature of 850 eV, and the signal-to-noise ratio is [approximately]20.