Study of the Power Distribution of Each Impedance in the Electrical Circuit of Ionized Gas Coagulation Equipment
Study of the Power Distribution of Each Impedance in the Electrical Circuit of Ionized Gas Coagulation Equipment
复制标题
电离气凝设备电路中各阻抗功率分布的研究
DOI:
10.1615/plasmamed.2016015853
复制
发表时间:
2015
期刊:
影响因子:
--
通讯作者:
Y. Ikehara
中科院分区:
文献类型:
--
作者:
H. Sakakita;S. Kiyama;Jaeho Kim;Hiromasa Yamada;I. Masukane;T. Niwa;Nobuyuki Shimizu;Y. Seto;M. Ichinose;Y. Ikehara
Medical plasma equipment is categorized as electrical equipment for use in medical treatment. The characteristics of both low-energy ionized gas coagulation equipment (LE-IGCE) as a low-temperature plasma source and high-energy ionized gas coagulation equipment (HE-IGCE) as a high-temperature plasma source strongly depend on the structure of the plasma equipment and the operating conditions. In this paper, to ensure the electrical safety of both the LE-IGCE and the HE-IGCE, the output current and power of both types of equipment are measured, and the power distribution is evaluated for each impedance in the electrical circuit. The power required to produce plasma using the LE-IGCE corresponds to the real power of the power supply, and it is higher than that required by the HE-IGCE. However, ohmic heating in the human body caused by the LE-IGCE is much lower than that caused by the HE-IGCE.
影响因子:
0.8
作者:
井上菜菜瀬;福本宗嗣;大槻均;入子英幸;SAKAKITA H. and IKEHARA Y
通讯作者:
SAKAKITA H. and IKEHARA Y