CONTRAST OPTIMIZATION FOR THE DETECTION OF FOCAL HEPATIC-LESIONS BY MR IMAGING AT 1.5-T

CONTRAST OPTIMIZATION FOR THE DETECTION OF FOCAL HEPATIC-LESIONS BY MR IMAGING AT 1.5-T
复制标题

DOI:
10.2214/ajr.149.6.1155
复制
发表时间:
1987-12-01
影响因子:
5
通讯作者:
HENDRICK, RE
HENDRICK, RE
中科院分区:
医学2区
文献类型:
--
作者:
FOLEY, WD;KNEELAND, JB;HENDRICK, RE

文献摘要

被引文献

相似文献

The relative efficacies of different spin-echo pulse sequences at 1.5 T were evaluated in the detection of focal hepatic disease. Pulse sequences compared were spin-echo with a repetition time (TR) of 200 msec and echo time (TE) of 20 msec, with six excitations; TR = 300 msec, TE = 20 msec, with 16 excitations (T1-weighted sequences); and a double spin-echo with TR = 2500 and Te = 25 and 70, with two excitations (proton-density-weighted and T2-weighted pulse sequences, respectively). Respiratory-motion compensation, which involved a reordering of the phase-encoding gradients (Exorcist), was used for the last two sequences. Spin-echo with TR = 2500 msec and TE = 70 msec was superior in lesion detection and contrast-to-noise ratio. The proton-density-weighted and T2-weighted sequences with respiratory compensation produced better artifact suppression than did the short TR, short TE T1-weighted sequence with temporal averaging. In contradistinction to prior results at 0.6 T, T2-weighted pulse sequences appear superior to T1-weighted pulse sequences with multiple excitations for both lesion detection and artifact suppression at 1.5 T.