Design and Performance of a TES X-ray Microcalorimeter Array for Energy Dispersive Spectroscopy on Scanning Transmission Electron Microscope
Design and Performance of a TES X-ray Microcalorimeter Array for Energy Dispersive Spectroscopy on Scanning Transmission Electron Microscope
复制标题
用于扫描透射电子显微镜能量色散光谱的 TES X 射线微量热仪阵列的设计和性能
DOI:
10.1007/s10909-016-1547-3
复制
发表时间:
2016
影响因子:
2
通讯作者:
T. Hara
中科院分区:
文献类型:
--
作者:
H. Muramatsu;K. Nagayoshi;T. Hayashi;K. Sakai;R. Yamamoto;K. Mitsuda;N. Yamasaki;K. Maehata;T. Hara
We discuss the design and performance of a transition edge sensor (TES) X-ray microcalorimeter array for scanning transmission electron microscope (STEM)–energy dispersive X-ray spectroscopy (EDS). The TES X-ray microcalorimeter has better energy resolution compared to conventional silicon drift detector and STEM–EDS utilizing a TES detector makes it possible to map the distribution of elements on a specimen in addition to analyze the composition. The requirement for a TES detector is a high counting rate (20 kcps), wide energy band (0.5–15 keV) and good energy resolution (10 eV) full width at half maximum. The major improvement of this development is to increase the maximum counting rate. In order to accommodate the high counting rate, we adopted anformat, 64-pixel array and common biasing scheme for the readout method. We did all design and fabrication of the device in house. With the device we have fabricated most recently, the pulse decay time is 40s which is expected to achieve 50 kcps. For a single pixel, the measured energy resolution was 7.8 eV at 5.9 keV. This device satisfies the requirements of counting rate and energy resolution, although several issues remain where the performance must be confirmed.