Evaluation of the Test-mate ChE (Cholinesterase) Field Kit in Acute Organophosphorus Poisoning
Evaluation of the Test-mate ChE (Cholinesterase) Field Kit in Acute Organophosphorus Poisoning
复制标题
DOI:
10.1016/j.annemergmed.2011.07.014
复制
发表时间:
2011-12-01
影响因子:
6.2
通讯作者:
Buckley, Nicholas A.
中科院分区:
文献类型:
--
作者:
Rajapakse, Bishan N.;Thiermann, Horst;Buckley, Nicholas A.
Study objective: Measurement of acetylcholinesterase (AChE) is recommended in the management of organophosphorus poisoning, which results in 200,000 deaths worldwide annually. The Test-mate ChE 400 is a portable field kit designed for detecting occupational organophosphorus exposure that measures RBC AChE and plasma cholinesterase (PChE) within 4 minutes. We evaluate Test-mate against a reference laboratory test in patients with acute organophosphorus self-poisoning.Methods: This was a cross-sectional comparison study of 14 patients with acute organophosphorus poisoning between May 2007 and June 2008. RBC AChE and PChE were measured in 96 and 91 samples, respectively, with the Test-mate ChE field kit and compared with a reference laboratory, using the limits of agreement method (Bland and Altman), kappa statistics, and Spearman's correlation coefficients.Results: There was good agreement between the Test-mate ChE and the reference laboratory for RBC AChE. The mean difference (Test-mate reference) was -0.62 U/g hemoglobin, 95% limits of agreement -10.84 to 9.59 U/g hemoglobin. Good agreement was also observed between the categories of mild, moderate, and severe RBC AChE inhibition (weighted kappa 0.85; 95% confidence interval [Cl] 0.83 to 0.87). Measurement of PChE also showed good agreement, with a mean difference (Test-mate reference) of +0.06 U/mL blood, 95% limits of agreement -0.41 to 0.53 U/mL blood. Spearman's correlation coefficients were 0.87 (95% Cl 0.81 to 0.91) for RBC AChE and 0.76 (95% Cl 0.66 to 0.84) for PChE. Analysis for within-subject correlation of subjects did not change the limits of agreement.Conclusion: The Test-mate ChE field kit reliably provides rapid measurement of RBC AChE in acute organophosphorus poisoning. [Ann Emerg Med. 2011;58:559-564.]