Surgeon Error in Performing Intraoperative Estimation of Stem Anteversion in Cementless Total Hip Arthroplasty
Surgeon Error in Performing Intraoperative Estimation of Stem Anteversion in Cementless Total Hip Arthroplasty
复制标题
DOI:
10.1016/j.arth.2013.03.006
复制
发表时间:
2013-10-01
影响因子:
3.5
通讯作者:
Iwamoto, Yukihide
中科院分区:
文献类型:
--
作者:
Hirata, Masanobu;Nakashima, Yasuharu;Iwamoto, Yukihide
To examine the accuracy of intraoperative estimation of stem anteversion in total hip arthroplasty (THA), we compared the intraoperatively estimated stem anteversion (estimated prosthetic anteversion) to stem anteversion measured by postoperative computed tomography (true anteversion) in 73 hips in 73 patients. Estimated prosthetic anteversion was significantly greater than true anteversion by 5.8 degrees, and the mean absolute value of surgeon error was 7.3 degrees ranging from 11 degrees underestimation to 25 degrees overestimation. Surgeons tended to overestimate when the true anteversion was smaller. A multivariate analysis showed that advanced knee osteoarthritis significantly increased surgeon error. These results indicated that estimated prosthetic anteversion was generally larger than true anteversion and that the grade of knee osteoarthritis affected the degree of surgeon error. (C) 2013 Elsevier Inc. All rights reserved.