Two-dimensional accelerated MP-RAGE imaging with flexible linear reordering
Two-dimensional accelerated MP-RAGE imaging with flexible linear reordering
复制标题
DOI:
10.1007/s10334-014-0430-y
复制
发表时间:
2014-10-01
影响因子:
2.3
通讯作者:
Stoecker, Tony
中科院分区:
文献类型:
--
作者:
Brenner, Daniel;Stirnberg, Ruediger;Stoecker, Tony
Implementation of an accelerated Magnetization Prepared RApid Gradient Echo (MP-RAGE) sequence for T1 weighted neuroimaging; exploiting modern MRI technologies to minimize scan time while preserving the image quality.A custom MP-RAGE sequence was implemented on a state-of-the-art 3T MR scanner equipped with a 32-channel receiver array head coil. The sequence utilized a shifted CAIPIRINHA k (y) -k (z) under-sampling pattern combined with elliptical scanning and a two-dimensional view ordering scheme to achieve high parallel imaging acceleration factors at maintained image contrast.It could be shown that MP-RAGE accelerated in two k-space directions outperforms single direction acceleration, which is the common practice with standard view ordering. Applying the CAIPIRINHA technique in conjunction with elliptical scanning further increased this benefit.By combining MP-RAGE with CAIPIRINHA sampling and elliptical scanning, the scan time can be reduced from 4-5 min to 2-3 min with insignificant reduction in image quality.