A pixel ionization chamber used as beam monitor at the Institut Curie- : Centre de Protontherapie de Orsay (CPO)
A pixel ionization chamber used as beam monitor at the Institut Curie- : Centre de Protontherapie de Orsay (CPO)
复制标题
DOI:
10.1016/j.nima.2006.06.024
复制
发表时间:
2006-09
影响因子:
1.4
通讯作者:
A. Rosa;M. A. Garella;F. Bourhaleb;R. Cirio;M. Donetti;S. Giordanengo;N. Givehchi;F. Marchetto;F. Martín;S. Meyroneinc;C. Peroni;G. Pittà
中科院分区:
文献类型:
--
作者:
A. Rosa;M. A. Garella;F. Bourhaleb;R. Cirio;M. Donetti;S. Giordanengo;N. Givehchi;F. Marchetto;F. Martín;S. Meyroneinc;C. Peroni;G. Pittà
The Dipartimento di Fisica Sperimentale and the Istituto Nazionale di Fisica Nucleare (INFN), Torino, in collaboration with the Institut Curie—Centre de Protontherapie de Orsay (CPO), have developed and built a pixel parallel plate ionization chamber to be used as monitor for the proton therapy beam line at the Institut Curie—CPO (Orsay, France). The sensitive area of the detector is (160×160)mm2, with the anode segmented in 1024 square pixels arranged in a 32×32 matrix; the area of each pixel is (5×5)mm2. The detector has been placed on the beam line just upstream of the last collimator to monitor the beam shape and to measure the stability and reproducibility of the delivery system. In this paper, we present a detailed description of the detector and the results of a set of preliminary tests.