Inaccuracies in MTS assays: major distorting effects of medium, serum albumin, and fatty acids
Inaccuracies in MTS assays: major distorting effects of medium, serum albumin, and fatty acids
复制标题
DOI:
10.2144/04373st05
复制
发表时间:
2004-09-01
期刊:
影响因子:
2.7
通讯作者:
Walker, AM
中科院分区:
文献类型:
--
作者:
Huang, KT;Chen, YH;Walker, AM
Soluble formazan assays are widely used for cell number assessment. However; in our hands, we observed frequent occasions in which the actual cell number was at odds with the assay reading. In this study, we have determined that (i) a large proportion of the reading obtained in commonly used culture media can be caused by media component amplification of formazan production in a way that cannot be corrected for by media-only controls; (ii) the albumin present in 10% serum can reduce the assay absorbance by 50% so that an actual doubling of cell number can be obscured; and (iii) this latter effect is dependent on the concentration of fatty acids. To counter these problems, we have developed a protocol that gives consistent readings that are fully representative of cell number while retaining some of the original advantages of soluble formazan assays.