A new X-ray computed tomography system for laboratory mouse imaging
A new X-ray computed tomography system for laboratory mouse imaging
复制标题
DOI:
10.1109/23.775579
复制
发表时间:
1999-06-01
影响因子:
1.8
通讯作者:
Allen, WC
中科院分区:
文献类型:
--
作者:
Paulus, MJ;Sari-Sarraf, H;Allen, WC
Two versions of a new high-resolution x-ray computed tomography system are being developed to screen mutagenized mice in the Oak Ridge National Laboratory Mammalian Genetics Research Facility. The first prototype employs a single-pixel cadmium zinc telluride detector with a pinhole collimator operating in pulse counting mode. The second version employs a phosphor screen/CCD detector operating in current mode. The major system hardware includes a low-energy x-ray tube, two linear translation stages and a rotational stage. For the single-pixel detector, image resolution is determined by the step size of the detector stage; preliminary images have been acquired at 100 Irm and 250 mu m resolutions. The resolution of the phosphor screen detector Is determined by the modulation transfer function of the phosphor screen; images with resolutions approaching 50 mu m have been acquired. The system performance with the two detectors is described and recent images are presented.