A laboratory testing and driving system for AIROF microelectrodes
A laboratory testing and driving system for AIROF microelectrodes
复制标题
AIROF微电极实验室测试和驱动系统
DOI:
--
复制
发表时间:
2004
期刊:
影响因子:
--
通讯作者:
S. Cogan
中科院分区:
文献类型:
--
作者:
N. R. Srivastava;P. Troyk;S. Cogan
The charge-injection currents of AIROF (activated iridium oxide film) microelectrodes, which are subjected to charge-balanced biphasic pulsing or monophasic current pulsing, have to be limited such that the anodic and cathodic voltage excursions are kept within safe limits of operation. In earlier studies it has been shown that when using anodic bias asymmetry in the magnitude of the balanced biphasic waveform can be used to increase the charge injection capacity of AIROF electrodes. We present the design of a single-channel testing and driving system for laboratory testing and driving of AIROF microelectrodes within safe charge-injection limits.