Effects of screw and host factors on insertion torque and pullout strength
Effects of screw and host factors on insertion torque and pullout strength
复制标题
DOI:
10.2319/070111-427.1
复制
发表时间:
2012-07-01
影响因子:
3.4
通讯作者:
Buschang, Peter H.
中科院分区:
文献类型:
--
作者:
Shah, Ankit H.;Behrents, Rolf G.;Buschang, Peter H.
Objective: To experimentally study the effects of altering implant length, outer diameter, cortical bone thickness, and cortical bone density on the primary stability of orthodontic miniscrew implants (MSIs).Materials and Methods: Maximum insertion torque (IT) and pullout strength (POS) of 216 MSIs were measured in synthetic bone with different cortical densities (0.64 g/cc or 0.55 g/cc) and cortical thicknesses (1 mm or 2 mm). Three MSIs were evaluated: 6-mm long/1.75-mm outer diameter,,3-mm long/1.75-mm outer diameter, and 3-mm long/2.0-mm outer diameter. To test POS, a vertical force was applied at the rate of 5 mm/min until failure occurred.Results: The 6-mm MSIs displayed significantly (P < .001) higher IT and POS than the 3-mm MSIs did. The 3-mm MSIs with 2.0-mm outer diameters showed significantly higher (P < .001) IT and POS than the 3-mm MSIs with 1.75-mm outer diameters. The IT and POS were significantly (P < .001) greater for the MSIs placed in thicker and denser cortical bone.Conclusion: Both outer diameter and length affect the stability of MSIs. Increases in cortical bone thickness and cortical bone density increase the primary stability of the MSIs. (Angle Orthod. 2012;82:603-610.)