HDR imaging evaluation of a NT-proBNP test with a mobile phone
HDR imaging evaluation of a NT-proBNP test with a mobile phone
复制标题
DOI:
10.1016/j.bios.2010.09.015
复制
发表时间:
2011-01-15
影响因子:
12.6
通讯作者:
Filippini, D.
中科院分区:
文献类型:
--
作者:
Preechaburana, P.;Macken, S.;Filippini, D.
The determination of NT-proBNP levels is key for the monitoring of patients with diagnosed heart failure and it is a routine measurement typically performed at health care centers, which would benefit from decentralized alternatives.Here we investigate the quantitative evaluation of a well-established NT-proBNP test using a standard mobile phone (Nokia 6720) as measuring platform rather than a dedicated instrument.A Java ME software developed for this application controls the illumination and imaging of the proBNP test under defined time intervals, which enables the composition of multi-exposure sets that are processed as high dynamic range (HDR) images for contrast enhancement.The results show that HDR processing significantly increases the sensitivity and resolution of the technique achieving a performance within the diagnostics range.These results demonstrate the feasibility to exploit a ubiquitous device to decentralize the evaluation of a routine test and identify key processing alternatives to bring the performance of such systems within the diagnostics range. (C) 2010 Elsevier B.V. All rights reserved.