High density dental materials and radiotherapy planning: Comparison of the dose predictions using superposition algorithm and fluence map Monte Carlo method with radiochromic film measurements
High density dental materials and radiotherapy planning: Comparison of the dose predictions using superposition algorithm and fluence map Monte Carlo method with radiochromic film measurements
复制标题
DOI:
10.1016/j.radonc.2006.10.010
复制
发表时间:
2006-12-01
影响因子:
5.7
通讯作者:
Thai, Van
中科院分区:
文献类型:
--
作者:
Spirydovich, Siarhei;Papiez, Lech;Thai, Van
Background and purpose: During radiotherapy planning high density dental materials create a major challenge in determining correct dose distribution inside patients with head-and-neck tumors.Patients and methods: In this work we investigated the absorbed dose distribution inside a solid water((R)) stab phantom with embedded high density material irradiated by a 6 MV photon beam of field size 10 x 10 cm. We evaluated the absorbed dose distribution with three different techniques: superposition algorithm, radiochromic film, and the fluence map Monte Carlo (FMMC) method.Results: The results obtained with radiochromic film and FMMC were in good agreement (within +/- 5% of the dose) with one another. The superposition algorithm, which is often considered superior to other commercially available dose calculation algorithms, produced appreciably less accurate results than FMMC. In particular, downstream from the high density cerrobend inhomogeneity the superposition algorithm predicts a higher dose than the measurement does by at least 10-16% depending upon the size of the inhomogeneity and the distance from it. Upstream of the high density inhomogeneities the superposition algorithm predicts a lower than measured dose due to its failure to predict the dose enhancement close to the inhomogeneity interface.Conclusions: The delivered dose downstream from a high density inhomogeneity would be significantly less than the prescribed dose calculated by the superposition algorithm. The FMMC method which is based on a hybrid of the superposition algorithm input fluence data and Monte Carlo can be a useful tool in predicting dose in the presence of high density (e.g. dental) materials. (c) 2006 Elsevier Ireland Ltd. All rights reserved.