Hydrodynamics-induced variability in the USP apparatus II dissolution test
Hydrodynamics-induced variability in the USP apparatus II dissolution test
复制标题
DOI:
10.1016/j.ijpharm.2004.08.003
复制
发表时间:
2005-03-23
影响因子:
5.8
通讯作者:
Muzzio, FJ
中科院分区:
文献类型:
--
作者:
Baxter, JL;Kukura, J;Muzzio, FJ
The USP tablet dissolution test is an analytical tool used for the verification of drug release processes and formulation selection within the pharmaceutical industry. Given the strong impact of this test, it is surprising that operating conditions and testing devices have been selected empirically. In fact, the flow phenomena in the USP test have received little attention in the past. An examination of the hydrodynamics in the USP apparatus II shows that the device is highly vulnerable to mixing problems that can affect testing performance and consistency. Experimental and computational techniques reveal that the flow field within the device is not uniform, and dissolution results can vary dramatically with the position of the tablet within the vessel. Specifically, computations predict sharp variations in the shear along the bottom of the vessel where the tablet is most likely to settle. Experiments in which the tablet location was carefully controlled reveal that the variation of shear within the testing device can affect the measured dissolution rate. (c) 2004 Published by Elsevier B.V.