Measurement of temperature-dependent specific heat of biological tissues

Measurement of temperature-dependent specific heat of biological tissues
复制标题

DOI:
10.1088/0967-3334/26/1/006
复制
发表时间:
2005-02-01
影响因子:
3.2
通讯作者:
Mahvi, DM
Mahvi, DM
中科院分区:
工程技术3区
文献类型:
--
作者:
Haemmerich, D;Schutt, DJ;Mahvi, DM

文献摘要

被引文献

相似文献

We measured specific heat directly by heating a sample uniformly between two electrodes by an electric generator. We minimized heat loss by styrofoam insulation. We measured temperature from multiple thermocouples at temperatures from 25 degreesC to 80 degreesC while heating the sample, and corrected for heat loss. We confirm method accuracy with a 2.5% agar-0.4% saline physical model and obtain specific heat of 4121 +/- 89 J (kg K)(-1), with an average error of 3.1%.