Sensitivity and Fading Characteristics of the 25 μm Pixel Size Imaging Plate for Transmission Electron Microscopes
Sensitivity and Fading Characteristics of the 25 μm Pixel Size Imaging Plate for Transmission Electron Microscopes
复制标题
25 μm像素尺寸透射电子显微镜成像板的灵敏度和衰落特性
DOI:
10.1093/oxfordjournals.jmicro.a023437
复制
发表时间:
1996
期刊:
影响因子:
--
通讯作者:
T. Oikawa
中科院分区:
文献类型:
--
作者:
A. Taniyama;D. Shindo;T. Oikawa
The sensitivity and fading characteristics of the new IP, which had a pixel size of 25 μm × 25 μm, were measured to calibrate the read-out intensity in quantitative analysis and to estimate the read-out efficiency of an IP-reader. The sensitivity showed good linearity in electron dose and a wide dynamic range of about five orders. However, the effective dynamic range was restricted to about four orders under one gain condition of the IP-reader. The fading characteristic of the new IP was independent of the electron intensity, and the decrease of the read-out intensity due to the fading was smaller than with the old IP which had a pixel size of 50 μm × 50 μm. Furthermore, the fading characteristic caused by repeated read-out was independent of the electron intensity, and the read-out intensity did not reach zero after read-out was repeated 10 times.